SHOWA SOKKI SHOCK METER (PEAK HOLD TYPE)
Code : 1340A
Brand : Showa Sokki
|
| Digital Accelerometer Model 1340A | Output cable(1m) | |
| Piezoelectric Pickup Model 2304A | Dry cell(6LR61) | |
| (with M6 screws,contact pin,handler) | Small-size slotted screwdriver | |
| Low noise cable LNC-1A-3(3m) | Shoulder case | |
| Magnet holder MG-1 | Operation Manual |

| (1)INPUT TRANSDUCER | Piezoelectric pickup Model 2304A Approx.50pC/m/s2 | |
| (2)INPUT TERMINAL | Charge input type,BNC connector | |
| (3)MEASURING RANGE | 1-1999m/s2 | |
| (Option:0.1-199.9m/s2(Peak)) | ||
| (4)RESOLUTION | 1m/s2(Peak) | |
| (Option:10m/s2(Peak)) | ||
| (5)FREQUENCY RANGE | 500Hz-5KHz,Within 5% | |
| (6)LOW-PASS FILTER | Cut-off frequency 500Hz(-3dB),-12dB/OCT | |
| (7)ACCURACY | 3% of reading,±1m/s2 | |
| (80Hz sine wave,100m/s2 vibration,23°C±3°C) | ||
| (8)DETECTION METHOD | VIBRATION Detection of absolute value of sine wave calibration, peak value indication. |
|
| SHOCK Detection of mean value,peak hold. | ||
| Releasable by reset button. | ||
| Peak hold is effective for acceleration of 10m/s2 | ||
| (For optional specifications,peak hold is effective for 100m/s2 | ||
| (9)RESET FUNCTION | For use in SHOCK MODE to clear held value. | |
| (10)AC OUTPUT | 100mv/(m/s2),±2V,output resistance 1KΩ,mini-plug. | |
| (Option:10mv/(m/s2) | ||
| (11)DIGITAL INDICATOR | LCD,3(1/2)digits | |
| Sample rate VIBRATION About 3 times/sec. | ||
| (12)POWER SUPPLY | 9V alkaline dry cell (6LR61)×1 Continuous use for more than 50 houres (Or 9V manganese cell (006P) ×1) |
|
| Bettery alarm about 7V or less,current consumption about 5mA. | ||
| (13)TEMPERATURE STABILITY | -10°C-50°C | |
| (14)WEIGHT | Body Approx. 220g | |
| Full set with accessory Approx. 800g |