MATSUZAWA Digital Rockwell Hardness Tester RMT-X Series
A pioneer of digital Rockwell hardness tester
Suitable for testing heat-treated samples and materials. Rockwell hardness tester of Matsuzawa continues pursuing usability anyone can easily operate Rockwell hardness tester of Matsuzawa. Loaded with a Color LCD display with backlight. The hardness tester secure very clear and high visibility. t makes acceptance and rejection judgement, conversion to other tabulation and computing for data memory(999 data), maximum value, variation, avarage, standard deviation etc.
Features
DIGITAL ROCKWELL RMT SERIES
Digital rockwell hardness tester RMT series
A pioneer of digital Rockwell hardness tester
Rockwell hardness tester of Matsuzawa continues pursuing usability anyone can easily operate Rockwell hardness tester of Matsuzawa
COLOR LCD TOUCH PANEL
Loaded with a Color LCD display with backlight. The hardness tester secure very clear and high visibility.
DISPLAY DEPTH
You can easily confirm depth on a measuring mode display.
Special Features
AUTO START
– After setting the reference load, the test load is applied automatically.
– The manual start is also possible
EASY SETTING
– Reference load can be easily set with a bar graph display on the bright screen and electronic sound.
– More over, an error-preventing device is also mounted.
PLASTIC MEASURING SYSTEM
– Specific measuring method for plastic in compliance with ASTM and JIS activated by one-touch easy operation.
– Time for reading the value the after unloading the test load can be set freely.
VARIOUS DATA PROCESSING FUNCTION
– It makes acceptance and rejection judgement, conversion to other tabulation and computing for data memory(999 data), maximum value, variation, average, standard deviation etc.
APPLICABLE TO CE & SAFETY
– Safety design applicable to EU low voltage command,EMC command and machine command. In addition to high-rigidity body.
– Overturn-preventive metal fittings are attached as standard to prevent overturing by earthquakes etc.
TRADITIONAL DIAL CHANGE-OVER SYSTEM
– Dial changeover systems make change the test load easily by turning the dial. Accordingly the design is dust-proof including the weights for test load.
Specifications
ITEM | RMT-3 | RMT-1 | ||||||
Preliminary Test Force | 29.42 98.07 N | 98.07 N | ||||||
3 10 kgf | 10 kgf | |||||||
Test Force | RMT-1 | 588.4 980.7 1471 N | ||||||
60 100 150 kgf | ||||||||
RMT-3 | 147.1 294.2 441.3 588.4 980.7 1471 N | |||||||
15 30 45 60 100 150 kgf | ||||||||
Preliminary Test Force Setting | LCD Graph and Electronic Signal Monitoring fine adjustment is unnecessary. | |||||||
Test Force Control | Automatic (Loading → Holding → Release) | |||||||
Start Operation | Automatic Start/Manual Start Selection | |||||||
Test Force Dwell Time | 3 ~ 99 sec | |||||||
Plastic Test Function | Standard function, time up to hardness value display after release: 3 ~ 99 sec | |||||||
Test Scale | 15N 30N 45N 15T 30T 45T 15W 30W 45W 15X 30X 45X 15Y 30Y 45Y | C D A G B F K E H P M L V S R | ||||||
C D A G B F K E H P M L V S R | ||||||||
Data Output | Centronics (Standard) | |||||||
RS232C or Serial Output selectable (Factory setting RS232C) | ||||||||
Data Memory | Max. 999 Test Data | |||||||
Hardness Conversion | Based on SAE J417B, ASTM E140 | |||||||
OK/NG Criteria | Upper and Lower Limit setting and HI/OK/LOW display | |||||||
Statistical Calculation | Batch conversion of memorized data. (Max. value, Min. value, dispersion (R), mean value) | |||||||
Converted Value, same too | ||||||||
RS232C Communication Line | Baud Rate: 1200/2400/4800/9600 bps selectable | |||||||
Error detection: Parity even/Parity none selectable | ||||||||
Accuracy | Conform to JIS B7726 and ASTM E-18 | |||||||
Max. Specimen Height | 200mm | |||||||
Max. Specimen Depth | 165mm | |||||||
Dimension | W220 x D540 x H830mm | |||||||
Weight | 100kg | |||||||
Power Supply | Single Phase AC100 ~ 240V 50/60Hz available for export models |