Specification

Model number: TI-56L

Calibration certificate: Included

Measurement range (mm): 1.5 – 250.0

Calibration certificate, certificate of test, traceability system diagram included

Application: Thick thing

Measuring method: Ultrasonic pulse reflection method

Reproducibility: +/-0.1mm

Measurement accuracy: 1.5 – 19.9mm/+/-0.1mm, 20.0 – 250.0mm/+/-0.5% rdg (rdg: The acronym for reading digits and means display value)

Used frequency: 5MHz

Minimum pipe size

Outer diameter: Approx φ30mm or more

Thickness: Approx 1.5mm or more

Probe: Cable length 1m, with LEMO connector, probe diameter φ13.0mm (contact surface diameter φ11.5mm)

Power supply: AA battery x 1 (included for test)

Size: 69 x 30 x 144mm

Weight (including cell): Body/approx 150g, Probe/50g

Accessories: Probe, couplant, 5mm thick standard piece for Calibration (attached to body), portable box

*One with calibration certificate includes calibration certificates (calibration certificate, test results report, traceability system) at the factory.

*A fee will be charged separately for issuing JCSS calibration.